- Place of Origin: China (Mainland)
- Model Number: WLD1C
It is a high performance spectrometer for routine analyses of aluminums, alnicos or non-ferrous metals, etc celerity.
SPECIFICATIONS:
Excitation SourceType High-energy low voltage spark
Power supply AC 220V±1%, 50Hz
Input power 1.2KVA
Capacitor 1μf, 2μf, 3μf, 7μf
Inductance L1=50μH L2=10μH
Optics UnitFocal length 0.75m
Grating Concave diffraction grating:2400L/mm
Dispersion 0.55nm/mm (1st order)
Wavelength range 220~550nm
Slit width Entrance slit 20μm
Exit slit 50/75μm
Maximum number of channel 36
Electrode holder sealed argon-filling holder
Readout UnitPhotometric stability RSD≤0.2%
Photometric mode partial integration
Data Processing Unit
Great PC
Operation of SoftwareSetting up working curve
Working cure standardization
Sample analyzing
Instrumental calibration
Self-testing
DIMENSIONS
Main unit 1700(L)×1000(D) ×1300(H)mm
Terminal 1300(L) ×700(D) ×1100(H)mm
Weight About 550Kg
CONFITIONS OF LAB
Condition Temperature 15~25
Relative Humidity ≤79%
Purity of argon 99.999%
Power supply Two Teams of Power Supply
Excitation Source AC 220V±10% 3KVA
Measuring System AC 220V±10% 1KVA
GND both Excitation Source and Measuring System have GND, at a distance of more than two meters, the resistance of GND is less than two ohm.
Floor Space more than fifteen m2